1. Zero Calibration
When adjusting the transmitter, please turn off the protection setting, which is usually the red switch in front of the meter and its status may be ON or OFF.

If it is ON, you will not be able to make the calibration adjustment.

Rosemount Differential Pressure Transmitter 1151DP, Type Supply:45 VDC MAX Output:4-20mA
Figure 1: Rosemount Transmitter 1151DP5E22M6Y1

Calibration Steps
First, ventilate the transmitter, and connect measuring circuit with the manual operator.

Then, operate on the manual operator as follows.
Calibration Steps
Finally, calibrate the transmitter through zero calibration according to the manual operator.

It is important that the zero calibration must be for the transmitter whose output is within 3% around zero.

2. DAC Calibration (Output Calibration)
The output calibration means to calibrate the signal output conversion from digital-to-analog (DAC).
Operate on the manual operator as follows.

DAC Calibration
And set the control loop as manual mode.

Before confirming, ...
... notice that if the positive and negative wiring is correct.
Finally, the output of the field device is set to 4mA, and the output of the transmitter is 4mA. Then record the actual value of the reference table, and input the value on the transmitter.

Check whether the output of the manual operator is consistent with the value of the reference table. If it is consistent, change the output to 20mA, and then repeat the operation until it's accurate. If not, you should continue to input the reference value and inspect whether it is consistent with the manual operator output.
Finally, set the control loop as automatic control.

Rosemount High Static Pressure Transmitter 1151HP Type Supply:45 VDC MAX Output:4-20mA
Figure 2: Rosemount Transmitter 1151HP7S22M4B1

3. ADC Calibration (Sensor Calibration)
The sensor calibration is to calibrate the analog-to-digital converter (ADC). The main work to be done is zero calibration and full calibration. Full calibration requires three times the accuracy of the transmitter. Operate on the manual operator as follows.
ADC Calibration
If the accuracy value of transmitter is 0.25mA, the lower limit output will be 0.06mA.

Related Info
Benefits of Rosemount Fieldbus Transmitter
Rosemount Transmitter Series 1151, 2088, 2051, 3051CA, 3051S

More About the Author

OKmarts.com Online Store | We provide leading brands of refrigeration compressors, servo motors, Air cylinder and more all over the world. More product's information please visit https://okmarts.com/

Total Views: 361Word Count: 325See All articles From Author

Add Comment

Automobiles Articles

1. Odac 24: Best Self Drive Car Rental Service In Panchkula
Author: Mahesh

2. From Protection To Personalization: The Modern Wheel Cover Revolution
Author: Creckk

3. Clearing Out The Old: Hassle-free Car Removal In Auckland
Author: Cars 4 Cash

4. Buy A Bobber And Enjoy Your Upcoming Ride
Author: Levi Fysh

5. Problems With A European Car You Must Be Certain Of
Author: Oliver Mertin

6. How To Ensure That You Pass Your Driving Test On Your First Attempt?
Author: Eleanor Thompson

7. Advantages Of Electric Light Commercial Vehicles For Sustainable Business Operations
Author: Ravi

8. Why Choose Cheap Tyres For Budget-friendly Choices?
Author: Juhi Kumari

9. Elevate Daily Living With Lift Chair Recliners
Author: Affordable Medical Equipment

10. Unleash Your Vehicle's True Power With Diesel Performance Upgrades
Author: Diesel Chip Tuning Pty Ltd

11. How To Troubleshoot And Fix Backhoe Loader Hydraulic System Parts
Author: Seetech Parts

12. Apa Itu Garansi Mobil Bekas?
Author: Bims Corner

13. Reliable Windscreen Services In South Auckland: Safety Starts With Clarity
Author: WindScreen

14. Stress Distribution And Load Management In Conventional Leaf Springs
Author: sonicoleafsprings

15. Your Event’s First Impression Starts At The Curb: Book The Right Limo In California
Author: Onyx Braun

Login To Account
Login Email:
Password:
Forgot Password?
New User?
Sign Up Newsletter
Email Address: